IN VIVO MULTIPHOTON FLUORESCENCE PHOTOBLEACHING RECOVERY (MPFPR) MEASUREMENTS
$0P41FY2002RRNIH
Cornell University Ithaca, Ithaca NY
Investigators
Linked publications & trials
Paper 26842836Paper 26224852Paper 23334294Paper 22770513Paper 22540300Paper 22042743Paper 21147979Paper 21118215Paper 20389554Paper 19905878Paper 18428347Paper 18163823Paper 18065276Paper 17728385Paper 16426080Paper 15851688Paper 15647000Paper 15208704Paper 15018559Paper 14517204Paper 12829512Paper 12775841Paper 12766225Paper 12717445Paper 12560545Paper 12324447Paper 12056830Paper 11964266Paper 11964263Paper 11922312Paper 11850419Paper 11832456Paper 11790090Paper 11578936Paper 11463636Paper 11371470Paper 11050231Paper 11034911Paper 10729332Paper 10664541Paper 10618386
Abstract
The major drawback of pulsed two-photon excitation schemes in fluorescence correlation spectroscopy compared to one-photon applications is the existence of comparably long dark periods (12ns) for pulsed excitation, since the fluorescence lifetime is for most fluorophors only several ns. By splitting the laser beam and using different delay lines, we built a pulse doubler and quadrupler that was applied to FCS. The signal-to-noise ratio could be improved by a factor of 1.5 with proper alignment.
View original record on NIH RePORTER →